| χb, χw | Dispersive shifts of the transmon qubit with respect to the buffer (b) and waste (w) resonators. Quoted as χ/2π in MHz. |
|---|---|
| κb, κw | Total linewidths (energy decay rates) of the buffer and waste resonators. Quoted as κ/2π in MHz. κw provides the irreversible dissipation that makes the click event irreversible. |
| |ξ| | Pump amplitude (dimensionless, in units of √photons) of the off-resonant pump tone that activates the four-wave-mixing process. Controls the parametric coupling g3 = 2|ξ|√(χbχw). |
| C | Cooperativity of the four-wave-mixing process, C = 4|ξ|2χbχw/(κbκw). Unit transfer efficiency is reached at C = 1. |
| η4wm | Transfer efficiency of a buffer photon into a qubit excitation via the four-wave-mixing process (Eq. 3). Peaks at 1 when C = 1. |
| T1 | Qubit energy-relaxation time (|e〉 → |g〉). Sets how long an absorbed excitation survives in the qubit. |
| Td | Detection window: duration during which the pump is on and an incoming photon can trigger a qubit excitation. |
| Tm | Measurement window: duration of the dispersive readout pulse on the waste resonator that checks the qubit state. |
| Tr | Reset window: conditional reset back to |g〉 (π-pulse + re-measurement if the qubit was found excited). Use the average reset time. |
| Tcycle | Td + Tm + Tr, the total duration of one detect–measure–reset cycle. |
| ηD | Duty-cycle efficiency = fraction of time the detector is actually listening, Td/Tcycle. |
| ηqubit | Probability that a qubit excitation created during detection survives until measurement, averaged over the detection window: (T1/Td)(1 − e−Td/T1). |
| ηRO | Single-shot readout fidelity for the excited state — probability of reading “click” given the qubit is in |e〉. |
| ηtheory | Total predicted operational efficiency: η4wm · ηRO · ηD · ηqubit. |
η4wm = 4C/(1+C)2, C = 4|ξ|2 χbχw/(κbκw). The 2π factors cancel in the ratio, so χ/2π and κ/2π in MHz may be entered directly.
ηD = Td/(Tm+Tr+Td), ηqubit = (T1/Td)(1 − e−Td/T1). Optimum: Tdopt ≈ √(2(Tm+Tr)T1) (Eq. G3).
ηtheory = η4wm · ηRO · ηD · ηqubit. Uses η4wm(|ξ|) from section 1 and ηD, ηqubit from section 2.
| Quantity | Expression |
|---|---|
| Cooperativity C | 4|ξ|2 χbχw/(κbκw) |
| 4WM efficiency η4wm | 4C/(1+C)2 (Eq. 3) |
| Optimal pump |ξ|opt | √(κbκw/(4χbχw)) (C=1) |
| Duty-cycle efficiency ηD | Td/(Tm+Tr+Td) |
| Qubit efficiency ηqubit | (T1/Td)(1 − e−Td/T1) |
| Optimal detection time | Tdopt ≈ √(2(Tm+Tr)T1) (Eq. G3) |
| Total efficiency | ηtheory = η4wm · ηRO · ηD · ηqubit |
Two kinds of measurement map onto the efficiencies on this page. Which one you look at determines which factors appear.
(a) End-of-detection qubit population pe (Fig. 2a-style). Read the qubit immediately after the detection window. In the weak-signal regime (〈nph〉 ≪ 1 per cycle):
pe(|ξ|) ≈ pebg + η4wm(|ξ|) · ηqubit · 〈nph〉
with 〈nph〉 = Φin · Td the mean number of signal photons during the detection window. Only η4wm and ηqubit enter — ηRO and ηD do not, because you are reading the population directly rather than converting to a per-second rate. Normalizing the curve to its peak gives the shape η4wm(|ξ|) independent of the absolute calibration of Φin.
(b) Click rate vs incoming photon rate (Fig. 3b-style — recommended). Record the detector click rate R for several input photon rates Φin and fit a line:
R = η · Φin + α
The linear fit extracts η and α simultaneously from one dataset, so dark counts are absorbed into the intercept and you do not need a separately calibrated “signal off” baseline (nor an accurate subtraction, which can drift between runs). For reliable numbers, sweep Φin over at least a decade and include a point or two deep enough that η · Φin ≫ α as well as a point near Φin = 0 to anchor the intercept.
The power sensitivity then follows from Eq. 1 of the paper: 𝒮 = ℏω√α / η.